• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number 03-2722-9
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/10/2017
Event Type  malfunction  
Manufacturer Narrative
Plant investigation is in process.A supplemental mdr will be submitted upon the completion of this activity.
 
Event Description
A user facility reported that a blood leak occurred approximately 1 ½ hours after initiation of the patient's hemodialysis (hd) treatment.The blood leak was noted as being a bloodline tubing leak.The machine did not alarm as it is not intended to do so.Blood was visually observed leaking externally from the red connection piece of the arterial tubing.No damage to the bloodline tubing was visible.Additionally, no damage to the bloodline packaging was observed.The patient¿s estimated blood loss (ebl) was noted as being approximately 300 cubic centimeters (cc).No patient adverse effects were experienced and no medical intervention was required as a result of this event.The patient completed treatment with a new set-up on the same machine.The complaint device was available to be returned to the manufacturer for evaluation.
 
Manufacturer Narrative
The device was returned to the manufacturer for physical evaluation.A visual examination of the returned device found that a gap was present between the main line tubing and the pump ¿t¿ connector; no other issues were detected on the tubing set.A records review was performed on the reported lot.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances or any associated rework during the manufacturing process which could be associated with the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The lot passed all release criteria.The investigation into the cause of the reported problem was able to confirm the failure mode.A visual examination found that a gap was present between the main line tubing and the pump ¿t¿ connector due to mis-assembled tubing.Therefore, the complaint has been deemed confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM COMBI SET
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX   88780
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key6602795
MDR Text Key76406667
Report Number8030665-2017-00287
Device Sequence Number1
Product Code FJK
UDI-Device Identifier00840861100293
UDI-Public00840861100293
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962081
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 06/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2019
Device Catalogue Number03-2722-9
Device Lot Number16SR01158
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2017
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received06/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/14/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age59 YR
Patient Weight99
-
-