• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS INC MERIT CUSTOM KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS INC MERIT CUSTOM KIT Back to Search Results
Catalog Number K12-MZT7201
Device Problems Delivered as Unsterile Product (1421); Unsealed Device Packaging (1444); Device Packaging Compromised (2916)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/16/2017
Event Type  malfunction  
Manufacturer Narrative
One device was returned for evaluation.The device was visually inspected and a defect in the sterile barrier was observed.The complaint is confirmed.The defect appears to have been caused by folded pouch material in the sealer.The device history record was reviewed and no exception documents were found.The complaint database was reviewed and no similar complaints for this lot number were found.
 
Event Description
The distributor reported a defect in the packaging.This was identified during their initial inspection of received product.The device was not sent to a user facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERIT CUSTOM KIT
Type of Device
CUSTOM KIT
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS INC
1600 merit parkway
south jordan UT 84095
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS INC
1600 merit parkway
south jordan UT 84095
Manufacturer Contact
casey hughes ms, cqe, csqp
1600 merit parkway
south jordan, UT 84095
8013164932
MDR Report Key6603451
MDR Text Key76389993
Report Number1721504-2017-00129
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K913682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 04/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Service and Testing Personnel
Device Expiration Date10/31/2019
Device Catalogue NumberK12-MZT7201
Device Lot NumberH1030823
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received05/24/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/10/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-