• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® I.C. ALL-SILICONE FOLEY CATHETER; FOLEY CATHETER (SILICONE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® I.C. ALL-SILICONE FOLEY CATHETER; FOLEY CATHETER (SILICONE) Back to Search Results
Catalog Number 1758SI14
Device Problem Mushroomed (2987)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/04/2017
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that a ridge was found on the catheter during inspection.
 
Manufacturer Narrative
Received 1 silicone catheter with the sample port connector still attached.The reported event was confirmed with an unknown cause.The preliminary visual inspection noted that there appeared to be a ridge/cuff on the balloon.A photo was obtained during the preliminary evaluation.However, the visual inspection, performed by medicon, noted no obvious defects and no cuff roll was observed.No other defects were observed.Per the functional evaluation, the balloon was inflated with air and deflated; cuff roll was not formed.After that, 10cc of a mix of water and blue methylene was introduced with a syringe and the catheter was left for 3 minutes resting on a flat surface.Then it was deflated by itself and cuff roll was not formed.Per the dimensional evaluation, the catheter active length was found within specification.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "caution: do not aspirate urine through drainage funnel wall.Note: aggressive traction, particularly in the presence of suturing, is not recommended for 100% silicone foley catheters.Recommended inflation capacities: 3cc balloon: use 5cc sterile water, 5cc balloon: use 10cc sterile water, 30cc balloon:use 35cc sterile water, do not exceed recommended capacities.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use." (b)(4).
 
Event Description
It was reported that a ridge was found on the catheter during inspection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® I.C. ALL-SILICONE FOLEY CATHETER
Type of Device
FOLEY CATHETER (SILICONE)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
amy gravley
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key6607066
MDR Text Key76593215
Report Number1018233-2017-02865
Device Sequence Number1
Product Code MJC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K984136
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 08/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/15/2021
Device Catalogue Number1758SI14
Device Lot NumberNGZL2339
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/17/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-