Our engineers have evaluated the returned device.Their investigation showed following: there was no endoprosthesis, deployment knob, deployment line, nor distal tip returned.The distal shaft, upon which the endoprosthesis was mounted, is approximately 10.7 with no distal tip at the end of the distal shaft.The distal end of the distal shaft showed no melting or disruption of the pebax coating.The length of the device from the tip of the distal shaft to the hub is approximately 71.5 cm.Based on the device examination performed, manufacturing anomalies were identified that potentially could have contributed to the event.
|