• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS 8000 PERFUSION SYSTEM; MONITOR AND/OR CONTROL, LEVEL SENSING, CARDIOPULMONARY BYPASS-SARNS 8000

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS 8000 PERFUSION SYSTEM; MONITOR AND/OR CONTROL, LEVEL SENSING, CARDIOPULMONARY BYPASS-SARNS 8000 Back to Search Results
Model Number 16413
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/10/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Evaluation is in progress, but not yet concluded.The fsr troubleshot the problem and determined the temperature/pressure printed circuit board (pcba) was at fault.The fsr replaced the temperature/pressure pcba.The unit operated to manufacturer's specifications.The temperature/pressure pcba was returned to the manufacturer for further evaluation.
 
Event Description
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, the channel a of arterial monitor would not display temperature.It would only display dashes.There was no patient involvement.
 
Manufacturer Narrative
(b)(4).The reported complaint was confirmed.During laboratory analysis, product surveillance technician (pst) observed that a damaged input connector caused the channel a input to not display temperature.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS 8000 PERFUSION SYSTEM
Type of Device
MONITOR AND/OR CONTROL, LEVEL SENSING, CARDIOPULMONARY BYPASS-SARNS 8000
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6609442
MDR Text Key76711929
Report Number1828100-2017-00258
Device Sequence Number1
Product Code DTW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K915183
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/02/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number16413
Device Catalogue Number16413
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received06/23/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-