Model Number 2AF283
Device Problem Air Leak (1008)
Patient Problem ST Segment Elevation (2059)
Event Date 05/12/2017
Event Type  malfunction  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, the consultant had bent the mapping catheter wire.The mapping catheter was replaced.Following the replacement of the mapping catheter, st elevation was observed.Additionally, it was indicated that air may have been introduced into the system.The st elevation resolved with no additional issues however the case was aborted.The patient was under general anesthesia.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, st elevation was observed.Additionally, it was indicated that air may have been introduced into the system.The st elevation resolved with no additional issues however the case was aborted.The patient was under general anesthesia.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the data files for the date of the event were returned and analyzed.The files showed fifteen injections with balloon catheter, 2af283 lot 73700.No system notices or issues were noted.The air ingress allegation could not be confirmed as no physical product was returned for analysis.Additionally, a known clinical issue was encountered.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER
Type of Device
PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6611850
MDR Text Key76728912
Report Number3002648230-2017-00256
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P100010/S015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 05/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2AF283
Device Catalogue Number2AF283
Device Lot Number22449
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/03/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age58 YR
-
-