During a mucosectomy, the subject device was used.During the procedure, the needle of the subject device was difficult to be removed and the physician could not perform the injection.The physician tried to retract the needle, but the needle could not be retracted into the sheath.The procedure was completed with a spare device of the same model.There was no patient injury reported.
|
This is a supplemental report for mfr report # 8010047-2017-00681 to provide additional information.The subject device has not been returned to olympus medical system corp.(omsc) for evaluation.The exact cause of the reported phenomenon could not be conclusively determined.The device history record for the lot indicated no abnormality with the event-related items below.1)needle extension/retraction 2)extended length of the needle 3)no bent, kink, crack and scratch on the tube.4)injection of the liquid based on the similar cases in the past, the needle might be unable to retract into the sheath due to the compressive buckling on the needle tube.The compressive buckling on the needle tube was likely caused by the large friction between the sheath and the needle tube when the needle was extended.It was likely that the friction between the outer tube and the needle increased by the following factors.1) the needle was extended/retracted while the tube was bent at inspection of operation.2) the slider was abruptly pushed.3) the buckling of the tube.The instruction manual of the device has already warned as follows; 1)when inserting the instrument into the endoscope, retract the needle into the sheath, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.2)insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.3)stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.
|