DEPUY SYNTHES POWER TOOLS OSCILL-SAW-ATTACHM F/532.001 532.010 532; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 532.021 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/11/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the locking knob seal was "incorrect" and the device failed the check the function of the quick saw blade coupling.Therefore, the reported condition was duplicated and confirmed.The assignable root cause was determined to be due to premature wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be sent accordingly.
|
|
Event Description
|
It was reported from (b)(6) that the top part of the oscillating saw attachment device dismantled.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|