• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI SPRINT QUATTRO SECURE; DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SPRINT QUATTRO SECURE; DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 6947M55
Device Problems Low impedance (2285); Device Contamination with Body Fluid (2317)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/05/2017
Event Type  malfunction  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the lead was attempted to be implanted but not used.The lead was found to have low lead impedance.The physician noticed blood lodged throughout the inner insulation of the lead body.The physician felt uncomfortable with the drop in impedance and the blood inside the lead body and decided to use a different lead.The lead was removed from implant and a new lead was implanted.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product event summary: the full lead was returned and analyzed.Analysis indicated that the inner insulation of the lead was extrinsically breached due to a cut.There was blood on the distal conductor of the lead and it was not obstructed.The outer insulation of the lead was extrinsically breached due to a cut.Visual summary analysis of the lead indicated damage at implant.The analyst confirmed, using destructive analysis, a cut through from the outer to the inner insulation and noted that blood was visible on distal conductor.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE
Type of Device
DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6627415
MDR Text Key77400343
Report Number2649622-2017-06748
Device Sequence Number1
Product Code NVY
UDI-Device Identifier00643169356610
UDI-Public00643169356610
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P920015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/02/2018
Device Model Number6947M55
Device Catalogue Number6947M55
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/05/2017
Initial Date FDA Received06/08/2017
Supplement Dates Manufacturer Received06/12/2017
06/12/2017
Supplement Dates FDA Received08/09/2017
10/03/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/15/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
-
-