• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER Back to Search Results
Catalog Number SGC0302
Device Problems Break (1069); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/16/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The customer reported the device was discarded.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
 
Event Description
This is filed to report the cable break.It was reported that this was a mitraclip procedure to treat mixed mitral regurgitation (mr) with a grade of 4.After successful preparation of the steerable guide catheter (sgc), following the instructions for use, just after insertion of the sgc into the femoral vein, the +/- knob was turned approximately 260 degrees, when a noise was heard.The sgc was removed and functions testing confirmed the cable had broken.A new sgc was used for the procedure and the successful deployment of 3 clips, reducing mr to 1.No adverse patient effects or clinically significant delay in the procedure were reported.No additional information was provided.
 
Manufacturer Narrative
(b)(4).The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no other incidents reported from this lot.The investigation was unable to determine a conclusive cause for the steerable guide catheter (sgc) cable break.The reported noise (device operates differently than expected) was likely a secondary effect of the cable break and therefore related to procedural conditions.Based on the information reviewed, there is no indication of a product quality issue with respect to design, manufacturing or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
STEERABLE GUIDE CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
MENLO PARK, CA REG# 3005070406
abbott vascular
26531 ynez rd.
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key6632238
MDR Text Key77405161
Report Number2024168-2017-04952
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/22/2018
Device Catalogue NumberSGC0302
Device Lot Number70221U121
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/16/2017
Initial Date FDA Received06/09/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/16/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-