Catalog Number 1010129-30 |
Device Problem
Tear, Rip or Hole in Device Packaging (2385)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/18/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that the outer packaging was so severly damaged that the sterile package was punctured; therefore, the sterility of the 8x30 mm acculink stent delivery system was compromised.The device was not used.There was no patient involvement.No additional information was provided.
|
|
Manufacturer Narrative
|
(b)(4).Visual inspections were performed on the returned device.The reported packaging damage was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation was unable to determine a conclusive cause for the reported packaging issue.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
|
|