While in use in the operating room, o-arm would not open from a closed position after use.O-arm was able to be moved to the foot of the bed and a c-arm was used to complete the case.Manufacturer response for o-arm, mct10 (per site reporter): at the time of the event support personnel from medtronics was contacted and attempted to assist the staff remotely.The patient was safely removed from the o-arm and a c-arm was used during the remainder of the surgical procedure.The technician completed service to the o-arm.The rotor was in the lateral position with door slightly open leading to door alignment pins being misaligned and engaging the dingus and preventing the door from opening further.The system was rehomed and adjusted upper door switch which was maladjusted leading to original issue.The system was checked out and unit operating as expected.
|