• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPECTRANETICS SPECTRANETICS LEAD LOCKING DEVICE; LLD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPECTRANETICS SPECTRANETICS LEAD LOCKING DEVICE; LLD Back to Search Results
Model Number 518-062
Device Problems Break (1069); Difficult to Remove (1528)
Patient Problem No Code Available (3191)
Event Date 05/17/2017
Event Type  Injury  
Manufacturer Narrative
Patient weight is unavailable.Device lot number, and expiration date are unavailable.Device manufacture date is dependent on the device lot number, thus is unavailable.
 
Event Description
It was reported that during a lead extraction procedure to remove 2 leads (ra and rv), the atrial baffle was damaged.Reportedly, this patient had previously undergone a mustard procedure to correct transposition of the great arteries.The physician was reportedly using a glidelight laser sheath to advance down the rv lead, with an lld device within the lead for traction.When the rv lead was being removed, the ra lead was found to be stuck to it.When the leads were pulled back, the atrial baffle was damaged.A bridge rescue balloon was deployed, which allowed the hole to clot off.No further intervention was required.
 
Manufacturer Narrative
Device 510k number has been corrected to reflect the most current and up to date number, as of the date of the initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Type of Device
LLD
Manufacturer (Section D)
SPECTRANETICS
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
719447-246
MDR Report Key6641962
MDR Text Key77604701
Report Number1721279-2017-00111
Device Sequence Number1
Product Code DRB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142116
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model Number518-062
Device Catalogue Number518-062
Device Lot NumberUNAVAILABLE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/19/2017
Initial Date FDA Received06/14/2017
Supplement Dates Manufacturer Received07/24/2017
Supplement Dates FDA Received08/09/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age40 YR
-
-