(b)(4).The anesthesia workstation was investigated on site by our field service engineer.The nozzle unit in the air gas module was replaced and the anesthesia workstation was returned to clinical use after successful function tests.The nozzle unit consists of a membrane, a mouthpiece and a feather spring.It is used for regulation of the gas flow through the gas module.The membrane in the nozzle unit is pressed against a mouthpiece with a feather spring to regulate the gas flow through the gas module.The received logs confirm the reported fault.The test log shows increased flow from the air gas module leading to increased pressure causing the pressure transducer test and the flow transducer test to fail.If the fault would occur during patient treatment, alarms such as airway pressure high and tidal/minute volume high are generated and a pressure relief via the fresh gas safety valve is done by the anesthesia system.The investigation of the replaced and returned nozzle unit could not reproduce the reported failing pressure transducer test.Several successful system check outs were performed.No visual defects explaining the reported event could be noticed.Based on the above information we are unable to determine the true cause of the reported event.
|