MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1712K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number MMT-1712K |
Device Problem
Erratic or Intermittent Display (1182)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/20/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|
|
Event Description
|
It was reported that the insulin pump had display anomaly.The screen was black with white flashing.Customer's blood glucose was unknown at the time of the incident.The insulin pump will not be returned for analysis.
|
|
Manufacturer Narrative
|
After battery installation the device gave an unexpected flashing white or blank display followed by an unexpected intermittent beep alarm due to cracked lcd controller.Unable to perform functional testing including self test, rewind, seating, basic occlusion, occlusion, force sensor and displacement test due to display anomaly.Unable to confirm missing segments or partial display due to display anomaly.
|
|