DENTSPLY CAULK PALODENT PLUS FORCEPS; INSTRUMENTS, DENTAL HAND Back to Search Results
Catalog Number 659810
Device Problem Break (1069)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
While no serious injury resulted in this event, there has been a previous report received where this malfunction resulted in a serious injury.Therefore, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it was reported that a pair of palodent plus forceps broke at the tip; no injury resulted.
 
Manufacturer Narrative
The actual device was not investigated because this failure is a known issue.A scar was initiated for tip breakage for this batch and investigation showed that the annealing process was interrupted when the electricity was shut down.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PALODENT PLUS FORCEPS
Type of Device
INSTRUMENTS, DENTAL HAND
Manufacturer (Section D)
DENTSPLY CAULK
38 w. clark ave.
milford DE 19963
Manufacturer (Section G)
DENTSPLY CAULK
38 w. clark ave.
milford DE 19963
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key6648890
MDR Text Key77834973
Report Number2515379-2017-00008
Device Sequence Number1
Product Code DZN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 09/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number659810
Device Lot NumberD0712
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/20/2017
Initial Date FDA Received06/16/2017
Supplement Dates Manufacturer Received08/25/2017
Supplement Dates FDA Received09/12/2017
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-