The investigation determined that higher than expected phyt results (26.58, 26.18 ug/ml versus expected 21.36 ug/ml) were obtained from biorad level 3 (lot 40903) quality control fluid using vitros phenytoin (phyt) reagent on a vitros 350 chemistry system.Biased results of the magnitude and direction observed may lead to inappropriate physician action if they were to occur undetected on patient samples.The unexpected vitros phyt results were generated from non-patient fluids, however the investigation cannot conclude that patient sample results were not affected and would not be affected if the event were to recur undetected.There was no allegation of patient harm as a result of the event.This report is number 1 of 2 mdr¿s for this event.Two (2) 3500a forms are being submitted for this event as 2 devices were involved.This report corresponds to ortho clinical diagnostics inc.Complaint number 1894687 / ivd 405843.
|