Model Number TF-25A
Device Problem Torn Material (3024)
Patient Problem Aortic Insufficiency (1715)
Event Date 05/29/2017
Event Type  Injury  
Event Description
On (b)(6) 2012, a 25 mm trifecta valve was implanted.On (b)(6) 2017, the valve was explanted due to aortic insufficiency.During explant, the non coronary leaflet of the valve was noted to be detached from the base of the valve.There were no signs of endocarditis.A 23mm carpentier edwards valve was implanted.The patient is reported to be stable.
 
Manufacturer Narrative
The results of this investigation concluded all three leaflets contained tears.There was circumferential fibrous pannus ingrowth on the inflow surface.There was no acute inflammation or significant calcifications observed in the valve.A review of the device history record showed the device met specifications prior to leaving abbott manufacturing facilities.There was no evidence found to suggest the cause of the tears and pannus were due to an intrinsic defect in the valve, as supported by review of the valve's device history record and the analysis performed.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM TRIFECTA VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA (CVD)
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA (CVD)
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6653142
MDR Text Key77969823
Report Number3008452825-2017-00130
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P100029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/29/2014
Device Model NumberTF-25A
Device Catalogue NumberTF-25A
Device Lot Number3621769
Other Device ID Number05414734052047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/31/2017
Initial Date FDA Received06/20/2017
Supplement Dates Manufacturer Received07/28/2017
Supplement Dates FDA Received08/03/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/31/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age84 YR
-
-