STRYKER ORTHOPAEDICS-MAHWAH EXETER V40 STEM 44MM NO 4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 0580-1-444
Device Problems Bent (1059); Material Deformation (2976)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 05/24/2017
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The customer reported that when the exeter stem was opened, the spigot protector appeared to be deformed.A new device was on hand to complete the case.
 
Manufacturer Narrative
An event regarding a damage issue involving a spigot protector for a exeter stem was reported.The event was confirmed.Device evaluation and results: a visual inspection completed by the supplier indicated: the yellow spigot shows a bent lug.We can see high material deformation on the bent lug, exactly located on the area where the stem inserter has to connect with the spigot.There is also deformation on the second lug.There are also marks of use with the instrument.Medical records received and evaluation: not performed as no adverse consequences to the patient were noted and no medical records were provided.Device history review: all devices in the reported lot were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusions: the investigation performed by the supplier on the returned part concluded "concluded that bent spigots are not due to manufacturing defect.The returned spigot shows bent lugs and visual marks of material bulge that are consistent with a misuse of the device".No further investigation is required at this time if further information becomes available this investigation will be re-opened.
 
Event Description
The surgeon noticed that the spigot protector was damage immediately when the exeter stem was opened; the spigot protector appeared to be deformed.A new device was on hand to complete the case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXETER V40 STEM 44MM NO 4
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
Manufacturer Contact
brian lauro
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6655485
MDR Text Key78143470
Report Number0002249697-2017-01953
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeGB
PMA/PMN Number
K121308
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 09/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0580-1-444
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/24/2017
Initial Date FDA Received06/20/2017
Supplement Dates Manufacturer Received08/18/2017
Supplement Dates FDA Received09/14/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-