• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY CAPTURA SERRATED FORCEPS WITH SPIKE; FCL, FORCEPS, BIOPSY, NON-ELECTRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY CAPTURA SERRATED FORCEPS WITH SPIKE; FCL, FORCEPS, BIOPSY, NON-ELECTRIC Back to Search Results
Catalog Number DBF-2.4SN-230SP-40-S
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Investigation evaluation: the product said to be involved was returned in an open pouch from the lot number provided in the report.The label matches the product returned.Our laboratory evaluation of the product said to be involved could not confirm the report as it was described.During the evaluation of the returned device, the device was functionally tested.When the handle of the device was manipulated, the forceps cups will open and close as intended.The device was then placed down an olympus gif-q20 (2.8 mm channel) endoscope.The endoscope was put in a torturous path.When the handle was manipulated the device would open and close as expected.The device was sent back to the supplier for evaluation.The supplier provided the following evaluation: one device from the reported event was returned in a zip type bag with proof of decontamination.The returned device was tested for "would not close".Upon functional testing and manipulation of the handle, the device opened and closed appropriately.The reported defect of "would not close" could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: the customer experienced issue of "would not close" was not confirmed.No corrective action will be implemented at this time.The instructions for use state the following in regard to product inspection: "beginning at the handle and moving toward the cups, uncoil the forceps making sure not to stretch the cable.Open and close the cups to verify smooth handle operation and appropriate cup action.Become familiar with the amount of handle movement required to operate the cups.If any irregularities are noted, do not use.Note: exercising the handle while the forceps is coiled may result in damage to the performance characteristics of the forceps." prior to distribution, all captura serrated forceps with spike are subjected to a visual inspection and functional test to ensure proper workability.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an esophagogastroduodenoscopy (egd), the physician used a cook captura serrated forceps with spike.The forceps did not close while down the endoscope inside the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURA SERRATED FORCEPS WITH SPIKE
Type of Device
FCL, FORCEPS, BIOPSY, NON-ELECTRIC
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key6658470
MDR Text Key78155490
Report Number1037905-2017-00419
Device Sequence Number1
Product Code FCL
UDI-Device Identifier00827002561335
UDI-Public(01)00827002561335(17)200106(10)W3814950
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 06/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberDBF-2.4SN-230SP-40-S
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location Hospital
Initial Date Manufacturer Received 05/23/2017
Initial Date FDA Received06/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-