The customer stated that they received an erroneous result for one patient sample tested for crep2 creatinine plus ver.2 (cre) on a cobas integra 400 plus (i400+).A sample from the patient was tested and resulted as 27 umol/l.This value was reported outside of the laboratory.A sample from the patient was tested on (b)(6) 2017, resulting as 165 umol/l.This sample was tested using a different reagent pack.It was asked, but it is not known if the 27 umol/l value and 165 umol/l value were from the same sample.The doctor said that the correct value for the patient is about 165 umol/l, because the patient suffers from a kidney disease.No adverse events were alleged to have occurred with the patient.The i400+ analyzer serial number was asked for, but not provided.
|
The customer mentioned that only one test was remaining in the cre reagent cassette when the low patient result was measured.The customer stated that this same situation occurred a few months ago, but provided no details.The customer assumed the issue to be related to the low volume in the reagent cassette.The reagent cassette was on board the analyzer for less than 4 weeks.A specific root cause could not be determined based on the provided information.Calibration, quality controls, and the performed check test were all good.Wash steps were programmed on the analyzer.Between the first and second measurement of the patient sample, there was a change in the reagent cassette.The low cre result was the last measurement performed out of the tests ordered on the sample.The system pipetted some other patient samples in between test measurements of the complained sample.No alarm was issued during the pipetting of the sample.The reaction curves for all other parameters tested on the patient sample appeared normal, only the cre result reaction curve was abnormal.It is assumed that some debris or something else was transferred into the reaction cell during measurement of the sample.It is also possible that sample material from other patient samples may have been transferred into the reaction cell as other patient samples were measured in between measurements of the complained sample.It is highly unlikely that only the last measurement out of a reagent cassette would cause an issue.The cassette contains enough dead volume of reagent to prevent this.
|