• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA SJM TRIFECTA VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA SJM TRIFECTA VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number TF-23A
Device Problem Torn Material (3024)
Patient Problem Aortic Insufficiency (1715)
Event Date 06/06/2017
Event Type  Injury  
Event Description
On (b)(6) 2014, a 23 mm trifecta valve was implanted.On (b)(6) 2017, aortic insufficiency was noted.The valve was explanted and a torn leaflet was confirmed.An unknown valve was implanted.The patient is reported to be stable.
 
Manufacturer Narrative
The results of this investigation concluded leaflets 2 and 3 were torn.There was circumferential fibrous pannus ingrowth on the inflow surface which resulted in the narrowing of the inflow diameter.Leaflet 3 contained a thin layer of outflow pannus.There was focal calcification in the inflow pannus of leaflet 3.No acute inflammation was present in the valve.A review of the device history record showed the device met specifications prior to leaving abbott manufacturing facilities.There was no evidence found to suggest the cause of the fibrin, tears, pannus, or calcification were due to an intrinsic defect in the valve, as supported by review of the valve's device history record and the analysis performed.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM TRIFECTA VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6666357
MDR Text Key78367692
Report Number3008452825-2017-00133
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P100029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/28/2016
Device Model NumberTF-23A
Device Catalogue NumberTF-23A
Device Lot Number4474332
Other Device ID Number05414734052030
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/06/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/29/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age73 YR
-
-