Device Problems Break (1069); Naturally Worn (2988)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/26/2017
Event Type  malfunction  
Manufacturer Narrative
The insulin pump had minor scratched lcd window, cracked case at display window corners, cracked battery tube threads, a test reservoir was installed and did not lock in place due to complete broken reservoir tube lip, missing o-ring (reservoir tube) and stained end cap sticker.
 
Event Description
The customer reported via phone call that their insulin pump was damaged.The customer's blood glucose level was unknown at the time of the incident.The customer stated that threads from the reservoir casing was worn out.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6668459
MDR Text Key78562139
Report Number3004209178-2017-76393
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
120010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 06/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-551NAL
Device Catalogue NumberMMT-551NAL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/01/2017
Initial Date FDA Received06/26/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age42 YR
Patient Weight71
-
-