• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. FEMORAL TRIAL SLAPHAMMER; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. FEMORAL TRIAL SLAPHAMMER; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLY Back to Search Results
Catalog Number 111140
Device Problems Break (1069); Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/01/2017
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.Review of the device history records indicates devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other similar events for the lot referenced.
 
Event Description
Threads broke off in trial.
 
Manufacturer Narrative
Correction (08/29/2017): manufacturer entity.An event regarding crack/fracture involving an mako other instrument was reported.The event was not confirmed.Method & results: device evaluation and results: a visual, functional and dimensional inspection could not be performed as the device was not returned.Medical records received and evaluation: no medical records or x-rays were made available for evaluation.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies complaint history review: there has been no other event for the lot referenced.Conclusions: the event could not be confirmed nor the root cause determined because the devices were not returned for evaluation and insufficient medical information was provided.If the devices and/or additional information are received, this investigation will be reopened and re-evaluated.Product surveillance will continue to monitor for trends.
 
Event Description
Threads broke off in trial.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL TRIAL SLAPHAMMER
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLY
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
timothy rice
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6675606
MDR Text Key78695115
Report Number3005738134-2017-00010
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number111140
Device Lot Number6040711
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/01/2017
Initial Date FDA Received06/28/2017
Supplement Dates Manufacturer Received08/29/2017
Supplement Dates FDA Received09/27/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/08/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-