MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AESTIVA 7100; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AESTIVA 7100; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Tidal Volume Fluctuations (1634)
Patient Problem No Patient Involvement (2645)
Event Date 06/05/2017
Event Type  malfunction  
Manufacturer Narrative
No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.A ge healthcare service representative performed a checkout of the equipment and confirmed the reported complaint.The flow sensor was replaced, and the unit was returned to service.
 
Event Description
The hospital reported the unit was over delivering tidal volume during pre-use checkout.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESTIVA 7100
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI
Manufacturer Contact
joseph seliga
3000 n grandview blvd.
waukesha, WI 
MDR Report Key6676617
MDR Text Key78681796
Report Number2112667-2017-01199
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K000706
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 06/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/20/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-