|
Model Number H749518120 |
Device Problems
Kinked (1339); Leak/Splash (1354)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Device evaluated by mfr: the device was returned for evaluation.Device analysis revealed a damaged in the lap joint.The telescope assembly was able to properly pull back, advance, and retract.Fluid was leaking from the lapjoint area when the catheter was flushed.No other visual damages or kinks were encountered during product analysis of the returned device.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
|
Event Description
|
Reportable based on investigation completed on 15jun2017.It was reported that sheath kinked and leaked occurred.An opticross¿ imaging catheter was used to view target lesion, there was no issue encountered during the first run.However, after removing the catheter from the patient during manual flushing with a syringe it was noted that there was no saline coming out from the tip end of the catheter, there was a leak at the kinked area mid distal of the opticross¿ imaging catheter.The procedure was completed with another opticross¿ imaging catheter.No patient complications were reported and the patient's status is stable.However, device analysis revealed a damage in the lapjoint area.
|
|
Search Alerts/Recalls
|
|
|