• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON GRIESHABER AG GRIESHABER REVOLUTION DSP SCISSORS; SCISSORS, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON GRIESHABER AG GRIESHABER REVOLUTION DSP SCISSORS; SCISSORS, OPHTHALMIC Back to Search Results
Catalog Number 705.52P
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/13/2017
Event Type  malfunction  
Manufacturer Narrative
No sample has been received by manufacturing for evaluation.No abnormalities that could have contributed to this event were found in the production documentation and the sample was released according to the manufacturer's acceptance criteria.A 100% final inspection is performed for this product.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Additional information has been requested.(b)(4).
 
Event Description
A nurse reported that an ophthalmic scissors device stopped closing during surgery.There was no impact to the patient.Additional information has been requested.
 
Manufacturer Narrative
The customer returned two curved scissor samples to manufacturing for evaluation.One sample was received with the original outer blister and cover foil.The other sample was in the original package, unopened.The opened sample was visually inspected with the aid of a photomicroscope and with various magnifications.It was found that the scissors is bent and one blade is broken off.In fact, it is unclear whether the sample condition is the same that the customer originally complained because the instrument was not protected during shipping and laid loosely in the blister.The other sample was tested and meets specification.The complaint history was reviewed two years back.It showed one comparable complaint with a bent scissor blade.It cannot exactly be determined how or when the damage occurred however, the damage to the complained sample is consistent with damage that can occur due to external force and or due to transport.No injuries have been reported or are expected related to this issue.A manufacturing or design related root cause for the damage of the complained device has not been identified.Possible root causes related to external impact which leads to this kind of damage could not be determined.The manufacturer has no influence on complaints which are in relation to external force.No further actions will be issued.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIESHABER REVOLUTION DSP SCISSORS
Type of Device
SCISSORS, OPHTHALMIC
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
nadia bailey
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key6680664
MDR Text Key79033930
Report Number3003398873-2017-00014
Device Sequence Number1
Product Code HNF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2019
Device Catalogue Number705.52P
Device Lot NumberF138820
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/13/2017
Initial Date FDA Received06/30/2017
Supplement Dates Manufacturer Received08/18/2017
Supplement Dates FDA Received08/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/21/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-