COV LP/LLC, CRYSTAL LAKE, MFG; SYRINGE, IRRIGATING (NON DENTAL) Back to Search Results
Model Number UNK NS
Device Problem Device Disinfection Or Sterilization Issue (2909)
Patient Problem Burn(s) (1757)
Event Date 03/01/2017
Event Type  Injury  
Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Customer reports: condensation was inside the syringe casing and when opened, the liquid has burned the skin of 3 individuals.
 
Manufacturer Narrative
A review of the device history record was unable to be performed without a lot number provided by the customer.A lot will not be released unless all acceptance criteria inspections per established sampling levels are within acceptable limits.In the event of a failure during the manufacturing process, the most probable root cause is insufficient mixing of vesta syde sq disinfectant and operator training for the controlled manufacturing cleanliness process.It is not clear from the information provided if the syringe was re-used.As indicated on the instructions for use provided, it is advised that syringes not be used if opened or damaged.A tamper evident seal is provided to determine if the syringe has been opened.Based on the information available and the investigation findings, a formal investigation is not deemed necessary at this time.In accordance with medical safety review, the severity for this incident is a level 1.At this point, a trend has not been established.If additional information is received warranting further analysis, the investigation will be resumed.This complaint will be used for tracking and trending purposes.A change order was completed to identify which department is responsible for vesta syde sq solution preparation and provide instructions on how to properly mix/dilute the cleaning agent and retraining for all affected employees was conducted as a corrective action.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Condensation was inside the syringe casing and when opened, the liquid has burned the skin of 3 individuals.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SYRINGE, IRRIGATING (NON DENTAL)
Manufacturer (Section D)
COV LP/LLC, CRYSTAL LAKE, MFG
815 tek dr
crystal lake IL 60014 8172
Manufacturer (Section G)
COV LP/LLC, CRYSTAL LAKE, MFG
815 tek dr
crystal lake IL 60014 8172
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key6682385
MDR Text Key78896946
Report Number1424643-2017-05011
Device Sequence Number1
Product Code KYZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberUNK NS
Device Catalogue NumberUNK NS
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/02/2017
Initial Date FDA Received06/30/2017
Supplement Dates Manufacturer Received03/02/2017
Supplement Dates FDA Received08/28/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-