MEDTRONIC SOFAMOR DANEK; CEMENT, BONE, VERTEBROPLASTY Back to Search Results
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
(revision surgery, persisting back pain) neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.
 
Event Description
It was reported in an abstract that the patient experienced old compression fracture in l1 in (b)(6) 2014.In bkp, bone cement was filled in cleft but not in cancellous bone.Postoperative, t2-weighted images revealed high signal in the whole circumference of the bone cement.Due to anterior migration of bone cement 3 weeks after bkp surgery and persistent low back pain with movement, anterior fusion by trans-retroperitoneal approach was carried out about 7 weeks after bkp.
 
Manufacturer Narrative
Additional information: literature citation: nobukiyo, m., aoki, m., ichimura, k., "the relationship between ¿pre- and postoperative mri findings¿ of the fractured vertebra and low back pain in patients having balloon kyphoplasty (bkp)".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK
1800 pyramid place
memphis TN 38132
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK
1800 pyramid place
memphis TN 38132
Manufacturer Contact
stacie ziemba
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key6683775
MDR Text Key78901226
Report Number1030489-2017-01648
Device Sequence Number1
Product Code NDN
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,l
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/03/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/05/2017
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age79 YR
-
-