Lead impedance demonstrated insufficient charge.The patient was seen in clinic on (b)(6)2017.While reviewing neurostimulator data, the clinician performed electrode impedance measurements and observed an insufficient charge impedance measurement on the 4th electrode of the lead connected to port 2.This was reconfirmed with multiple impedance measurements.An x-ray of the patient's head was performed which showed that the lead in port 2 was not fully seated in the connector cover.The patient had a consult with the neurosurgeon on (b)(6)2017 and was scheduled for a revision procedure on (b)(6)2017.During the surgical procedure, the connector cover was replaced and the lead in port 2 was re-seated.Normal impedance measurements were documented post-procedure.No indication of product failure was reported.
|