• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. FAN SPRAY KIT; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. FAN SPRAY KIT; LAVAGE, JET Back to Search Results
Model Number N/A
Device Problem Expulsion (2933)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/10/2017
Event Type  malfunction  
Manufacturer Narrative
The complaint is being reported by zimmer biomet as (b)(4).The customers reported event was that the device did not work and is considered a complaint out of the box (coob).Review of the device history record for (b)(4), lot number 63452275, identified no deviations or anomalies.Product examination found that the battery pack had ruptured from overheating batteries.The sealed packaging had not been opened.This complaint is confirmed.The reported event claimed that the sealed unit had ruptured before it had been opened.While rare, this kind of event occurs when the batteries overheat from a short circuit.To address this issue, change notice has been implemented to adjust the length of the wires inside the battery pack.This means that there will no longer be a need to tightly fold wires inside the battery pack in order for them to reach their respective circuit contacts.This reduces the risk of a short circuit.However, the root cause cannot be specifically determined with the provided information.
 
Event Description
It was reported that the battery pack of the device burst in the package before it had been opened and used.No adverse events have been reported as a result of the malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAN SPRAY KIT
Type of Device
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key6691236
MDR Text Key79237152
Report Number0001526350-2017-00405
Device Sequence Number1
Product Code FQH
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Physician
Type of Report Initial
Report Date 07/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2019
Device Model NumberN/A
Device Catalogue Number00515047500
Device Lot Number63452275
Other Device ID Number00889024375154
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/05/2017
Initial Date FDA Received07/06/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/16/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-