Model Number 16413
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/16/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Evaluation is in progress, but not yet concluded.Per field service representative (fsr), the arterial monitor functions to specification.The fsr could not duplicate the complaint.A replacement monitor was installed.The unit is operating within manufacturer's specifications.Unit was returned to manufacturer for additional evaluation.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the display for arterial pressure was not showing any information.They used cardioplegia monitor channel b in place of arterial pressure for this case.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
During laboratory analysis, the product surveillance technician (pst) connected the monitor to power supply and connected a pressure simulator to the input of the monitor and the monitor operated for eight hours with no failure of the monitor.The pst soaked the monitor at 10°c for three hours and when removed it displayed pressure.The pst operated the monitor for 72 hours with no failure on the monitor to display pressure.The pst agitated the internal connections and circuit boards with no failure of the monitor to display pressure.
 
Manufacturer Narrative
The reported complaint was not confirmed.The field service representative (fsr) replaced the power board in the safety monitor.This board supplies power to the arterial monitor and may have an intermittent problem.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS 8000 PERFUSION SYSTEM
Type of Device
MONITOR AND/OR CONTROL, LEVEL SENSING, CARDIOPULMONARY BYPASS-SARNS 8000
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6700494
MDR Text Key80054242
Report Number1828100-2017-00300
Device Sequence Number1
Product Code DTW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K915183
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 10/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16413
Device Catalogue Number16413
Other Device ID NumberGTIN (01)00886799000205
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received09/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/14/1994
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-