• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problems Loose or Intermittent Connection (1371); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/15/2017
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of monitor sn (b)(4) has been completed.The reported problem (code 204 / loose connection) was confirmed.As received, the monitor displayed code 204 and the belt receptacle was broken off of the monitor case, damaging internal wires.The cause of the code 204 is a disruption in communication between the monitor and electrode belt.The cause of the disruption in communication is the damaged receptacle and wires.The root cause of the damaged receptacle and wires is excessive force placed at the receptacle.No adverse event resulted from the damaged monitor.
 
Event Description
A us distributor contacted zoll to report that a patient's device was producing service code 204 and that the belt connection was loose.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
katelynn mains
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key6702940
MDR Text Key79897684
Report Number3008642652-2017-05842
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005005
UDI-Public00855778005005
Combination Product (y/n)N
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial
Report Date 07/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2017
Was the Report Sent to FDA? No
Date Manufacturer Received06/29/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/21/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-