Device evaluation of monitor sn (b)(4) has been completed.The reported problem (code 204 / loose connection) was confirmed.As received, the monitor displayed code 204 and the belt receptacle was broken off of the monitor case, damaging internal wires.The cause of the code 204 is a disruption in communication between the monitor and electrode belt.The cause of the disruption in communication is the damaged receptacle and wires.The root cause of the damaged receptacle and wires is excessive force placed at the receptacle.No adverse event resulted from the damaged monitor.
|