ORTHO-CLINICAL DIAGNOSTICS VITROS CHEMISTRY PRODUCTS GLU SLIDES; IN-VITRO DIAGNOSTIC Back to Search Results
Catalog Number 1707801
Device Problem High Readings (2459)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/13/2017
Event Type  malfunction  
Manufacturer Narrative
The investigation determined the higher than expected vitros glu results were obtained for multiple quality control fluids and patient samples run on a vitros 5600 integrated system.The assignable cause of the event is unknown, however, the most likely assignable cause is a non-reactive cartridge caused by having the integrity of the foil wrapper compromised and exposing the slides to ambient conditions, although this could not be confirmed.For the vitros glu assay, a non-reactive cartridge causes higher than expected results.Acceptable glu results were obtained when the same samples were tested with an alternate vitros glu slide lot 0030-0918-2862 cartridge, and quality control results processed on alternate vitros glu slide lot 0030-0918-2862 cartridges prior to, and after the event were acceptable.This would indicate the event was isolated to the affected vitros glu slide lot 0030-0918-2862 cartridge in use at the time of the event.
 
Event Description
The customer obtained higher than expected vitros glu results for multiple quality control fluids and patient samples run on a vitros 5600 integrated system.(b)(6).Biased results of the direction and magnitude observed may lead to inappropriate physician action.The affected results were reported outside the laboratory; however, a corrected report was issued once repeat testing was complete.There was no allegation of patient harm as a result of this event.This report is number 22 of 33 mdr's for this event.Thirty-three (33) 3500a forms are being submitted for this event as 33 devices were involved.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITROS CHEMISTRY PRODUCTS GLU SLIDES
Type of Device
IN-VITRO DIAGNOSTIC
Manufacturer (Section D)
ORTHO-CLINICAL DIAGNOSTICS
100 indigo creek drive
rochester NY 14626
Manufacturer (Section G)
ORTHO-CLINICAL DIAGNOSTICS
513 technology blvd.
rochester NY 14652
Manufacturer Contact
james a stevens
100 indigo creek drive
rochester, NY 14626
5854533000
MDR Report Key6710021
MDR Text Key80282458
Report Number1319809-2017-00120
Device Sequence Number1
Product Code CGA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Medical Technologist
Type of Report Initial
Report Date 07/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/01/2017
Device Catalogue Number1707801
Device Lot Number0030-0918-2862
Other Device ID Number10758750009572
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/14/2017
Initial Date FDA Received07/13/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/06/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-