• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS HEMOSTAR LONG-TERM HEMODIALYSIS CATHETER 14.5F STANDARD KIT (STRAIGHT) (19CM); BLOOD ACCESS AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS HEMOSTAR LONG-TERM HEMODIALYSIS CATHETER 14.5F STANDARD KIT (STRAIGHT) (19CM); BLOOD ACCESS AND ACCESSORIES Back to Search Results
Model Number N/A
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 06/19/2017
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of rebp0259 showed no other similar product complaint(s) from this lot number.
 
Event Description
It was reported that foreign material was found inside the sealed package.The device was not used on a patient.
 
Manufacturer Narrative
The complaint of a hair in the packaging is confirmed, manufacturing related.The product returned is one sealed hgstar 14.5 fr kit.A hair is observed within the outer packaging.The seals and packaging appear undamaged and sealed.As the package is sealed and the hair is visible through the packaging, the complaint is confirmed, manufacturing related.The lot history review for rebp0259 revealed no similar incidents to date.Corrective actions will be implemented at the manufacturing facility to prevent recurrence.
 
Event Description
It was reported that foreign material was found inside the sealed package.The device was not used on a patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOSTAR LONG-TERM HEMODIALYSIS CATHETER 14.5F STANDARD KIT (STRAIGHT) (19CM)
Type of Device
BLOOD ACCESS AND ACCESSORIES
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas
MX  
Manufacturer Contact
kelsey erickson
605 n. 5600 w.
salt lake city, UT 84116
8015225937
MDR Report Key6711218
MDR Text Key80054958
Report Number3006260740-2017-01011
Device Sequence Number1
Product Code MSD
UDI-Device Identifier00801741013386
UDI-Public(01)00801741013386
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K051748
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 09/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2018
Device Model NumberN/A
Device Catalogue Number5833690
Device Lot NumberREBP0259
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2017
Is the Reporter a Health Professional? Yes
Event Location Hospital
Date Manufacturer Received07/26/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-