• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND DYNAGEN; IMPLANTABLE CHF GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND DYNAGEN; IMPLANTABLE CHF GENERATOR Back to Search Results
Model Number G158
Device Problems Pacing Problem (1439); Device Operates Differently Than Expected (2913)
Patient Problem Atrial Tachycardia (1731)
Event Date 05/15/2017
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy defibrillator (crt-d) was reported to have an unspecified device failure.Boston scientific technical services (ts) reviewed presenting electrogram (egm) in which atrial sensing, biventricular pace were observed and also noted that the left and right ventricular leads were separated with an interventricular delay (vvd).Pacemaker mediated tachycardia (pmt) was observed which did not terminate the rhythm as the atrial rate continued to marched along at 130 beats per minute (bpm).Attempts to obtain additional information from the field were unsuccessful.This crt- d remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key6718630
MDR Text Key80203870
Report Number2124215-2017-11597
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534904
UDI-Public(01)00802526534904(17)20180729
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/29/2018
Device Model NumberG158
Other Device ID NumberDYNAGEN X4 CRT-D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received05/16/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/29/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0292; 4671; 4674; 7740; G158
Patient Age63 YR
-
-