• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALERE SAN DIEGO, INC. CONSULT HCG URIN/SRM COMBO 5002 25; HCG PREGNANCY TEST

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALERE SAN DIEGO, INC. CONSULT HCG URIN/SRM COMBO 5002 25; HCG PREGNANCY TEST Back to Search Results
Model Number FHC-A202
Device Problem False Positive Result (1227)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/12/2017
Event Type  Injury  
Manufacturer Narrative
Investigation conclusion: the customer's observation was not replicated in-house with retention and return products.Retention and return devices were tested with in-house clinical hcg negative urine samples as well as low concentration hcg standards.All hcg results were negative at read time.No false positive results were obtained during in-house testing.Manufacturing batch record review did not uncover any abnormalities.Based on the information available, there is no indication of a product deficiency and no corrective action is required.
 
Event Description
The customer reported the following events occurring for one patient: on (b)(6) 2017, the patient was tested prior to a scheduled iud insertion procedure.A urine sample was collected and tested using two consult hcg urine/serum combo tests.Each consult hcg urine/serum combo test produced a positive result.The procedure was delayed and a serum sample was drawn and sent out for testing.The serum beta quantitative test produced a negative hcg result of <2 miu/ml.After receiving the laboratory result, the iud insertion procedure was rescheduled.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSULT HCG URIN/SRM COMBO 5002 25
Type of Device
HCG PREGNANCY TEST
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9975 summers ridge road
san diego CA 92121
Manufacturer Contact
ya-ling king
9975 summers ridge road
san diego, CA 92121
8588052084
MDR Report Key6719754
MDR Text Key80247698
Report Number2027969-2017-00116
Device Sequence Number1
Product Code JHI
UDI-Device Identifier20612479202768
UDI-Public(01)20612479202768(17)180731(10)HCG6090154
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFHC-A202
Device Lot NumberHCG6090154
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/19/2017
Initial Date FDA Received07/17/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-