It was reported that during a laparoscopic cholecystectomy procedure, ¿first attempt at firing, no clip.Second attempt showed distorted bent tip and malformed clip.The device was discarded and new device opened.There was a three minute delay to procedure.There was no adverse event to patient.¿.
|
(b)(4).Batch # p91h4h.The analysis results found that the el5ml device was returned with no damage in the external components; upon visual inspection, two clips were noted to be jammed inside the shaft.The clips were removed in order to perform functional testing.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 8 conforming clips and the lockout mechanism was found to be non-functional.In order to evaluate the device¿s internal components, the instrument was disassembled.Upon disassembling of the device, the ratchet pawl was found damage causing the lockout failures.The jammed clip may have caused the device to not fire the clips properly or at all; however, cause not determined could be reached as to what may have caused the clip jamming.No conclusion could be reached as to what may have caused the reported incident.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|