• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. CUSTOM COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number 03-2722-9
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/03/2017
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event Description
A user facility nurse reported that during a patient¿s hemodialysis (hd) treatment, approximately 1 ml (milliliters) of blood leaked from the venous port of the fresenius bloodline.Follow-up information with the clinic manager revealed that approximately five to ten minutes after the initiation of hd treatment, the patient care technician (pct) noticed a significant amount of air on the header of the dialyzer.The pct confirmed that all connections were secure and tight.It was then noticed that there was some blood coming out of the venous blood sample port.There was no machine alarm and no reported machine malfunction.The patient's estimated blood loss (ebl) was noted as being approximately 1ml.The treatment was immediately stopped and the patient was set-up with new supplies.No patient adverse effects or injuries were experienced and no medical intervention was required.The patient was able to complete treatment on the hd machine without any further interruptions or complications.The bloodline device was not available to be returned to the manufacturer as it was discarded.
 
Manufacturer Narrative
Plant investigation: the reported complaint was not confirmed as the complaint device was not available for manufacturer evaluation.As such, a definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.A records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.In addition, the dhr review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM COMBI SET
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX   88780
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key6727563
MDR Text Key80513906
Report Number8030665-2017-00458
Device Sequence Number1
Product Code FJK
UDI-Device Identifier00840861100293
UDI-Public00840861100293
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962081
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 08/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2020
Device Catalogue Number03-2722-9
Device Lot Number17CR01418
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received07/25/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/27/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age87 YR
Patient Weight83
-
-