Model Number CD3231-40
Device Problems Premature Discharge of Battery (1057); Failure to Interrogate (1332); Telemetry Discrepancy (1629)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/06/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the implantable cardioverter defibrillator exhibited premature battery depletion remotely via transmitter on (b)(6) 2017.The asymptomatic patient presented in the clinic on (b)(6) 2017, the device exhibited communication anomaly and could not be interrogated.It was suggested the device should be changed out.
 
Manufacturer Narrative
(b)(4).
 
Event Description
New information on 08/08 stated that the device was explanted and replaced, the patient was stable throughout the whole procedure.
 
Manufacturer Narrative
Premature battery depletion was confirmed by analysis.Bench testing on the device was performed, and no sources of high current were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY CRT-D
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6728288
MDR Text Key80549070
Report Number2017865-2017-06667
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 01/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2014
Device Model NumberCD3231-40
Device Catalogue NumberCD3231-40
Device Lot Number3872276
Other Device ID Number05414734504362
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/17/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/27/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0116-2017
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age69 YR
-
-