• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ASSURA CRT-D QUAD RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ASSURA CRT-D QUAD RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3365-40Q
Device Problems Over-Sensing (1438); Pacing Inadequately (1442)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/27/2017
Event Type  malfunction  
Event Description
It was reported that a patient presented via merlin.Net transmission.The right ventricular lead showed multiple episodes of oversensing.Some deflections were also seen in the signal on the discrimination channel.Follow up attempts have been made with the patient but has been unsuccessful.No intervention has been performed.No patient symptoms were reported.
 
Event Description
The implantable cardioverter defibrillator exhibited oversensing causing myopotential inhibition.The device was reprogrammed to resolve the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA CRT-D QUAD RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6728821
MDR Text Key80565568
Report Number2938836-2017-30559
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2016
Device Model NumberCD3365-40Q
Device Catalogue NumberCD3365-40Q
Device Lot Number4449178
Other Device ID Number05414734508292
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/10/2017
Initial Date FDA Received07/20/2017
Supplement Dates Manufacturer Received07/21/2017
Supplement Dates FDA Received07/27/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age81 YR
-
-