• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION TERUMO HX2 TEMPERATURE MANAGEMENT SYSTEM; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-HX2 TEMPERATURE MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION TERUMO HX2 TEMPERATURE MANAGEMENT SYSTEM; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-HX2 TEMPERATURE MANAGEMENT Back to Search Results
Model Number 809810
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/28/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the bottom of the unit was leaking water.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The field service representative (fsr) found channel ''a'' outlet water fitting was cracked and leaking.He replaced fitting and tested operation.The unit operated to the manufacturer's specifications.During laboratory analysis, the product surveillance technician (pst) found the fitting to be broken.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO HX2 TEMPERATURE MANAGEMENT SYSTEM
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-HX2 TEMPERATURE MANAGEMENT
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6729978
MDR Text Key80700431
Report Number1828100-2017-00318
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number809810
Device Catalogue Number809810
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/06/2017
Initial Date FDA Received07/20/2017
Supplement Dates Manufacturer Received07/24/2017
Supplement Dates FDA Received08/15/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/11/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-