Model Number CD3357-40Q
Device Problems Over-Sensing (1438); Pacing Inadequately (1442)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/25/2017
Event Type  malfunction  
Event Description
It was reported that a patient presented via merlin.Net transmission.Upon reviewing the electrogram, the implantable cardioverter defibrillator exhibited noise causing myopotential inhibition.No intervention has been performed but reprogramming has been scheduled later.No patient symptoms have been reported and the patient will be monitored with routine follow-up.
 
Event Description
New information received: patient presented in clinic due to reoccurrence of the oversensing due to myopotential oversensing.The oversensing issue was not reproducible in the clinic.New programming changes were recommended.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6730898
MDR Text Key80622572
Report Number2938836-2017-30574
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2018
Device Model NumberCD3357-40Q
Device Catalogue NumberCD3357-40Q
Device Lot NumberA000032737
Other Device ID Number05414734508216
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/19/2017
Initial Date FDA Received07/21/2017
Supplement Dates Manufacturer Received08/23/2017
Supplement Dates FDA Received09/18/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/05/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
-
-