SYNTHES BETTLACH 2.0MM IMF SCREW SELF-DRILLING 12MM; SCREW FIXATION INTRAOSSEOUS Back to Search Results
Catalog Number 201.932
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/04/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device broke intra-operatively and was not implanted/explanted.Without a lot number the device history records review could not be completed.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes europe reports an event in (b)(6) as follows: it was reported that during a surgery, two imf screws broke.This is report 2 of 2 for (b)(4).
 
Manufacturer Narrative
Product development investigation was completed; the report indicates that: we have received two imf screws for investigation.The visual inspection of both screws has shown that both tips are broken off.The broken off parts were not returned for evaluation.The microscopic investigation has shown damages on the cross slot.Unfortunately we are not able to determine the exact cause which has led to this occurrence, due to the damages on the cross slot we can only assume that a mechanical overload situation has led to the breakage.The device history record could not be reviewed, as the lot numbers of both articles are unknown.Because of the damage, the complaint relevant dimensions cannot be checked for dimensional accuracy of the valid manufacturing specifications device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0MM IMF SCREW SELF-DRILLING 12MM
Type of Device
SCREW FIXATION INTRAOSSEOUS
Manufacturer (Section D)
SYNTHES BETTLACH
muracherstrasse 3
bettlach CH254 4
SZ  CH2544
Manufacturer (Section G)
SYNTHES BETTLACH
muracherstrasse 3
bettlach CH254 4
SZ   CH2544
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6732077
MDR Text Key80914434
Report Number9612488-2017-10315
Device Sequence Number1
Product Code DZL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K010527
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 06/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number201.932
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/11/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-