DEPUY ORTHOPAEDICS, INC. 1818910 UNKNOWN DEPUY XL TAPER SLEEVE ADAPTOR; HIP FEMORAL STEM/SLEEVE Back to Search Results
Catalog Number UNK-HIP
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problems Host-Tissue Reaction (1297); Muscle Weakness (1967); Pain (1994); Cramp(s) (2193)
Event Date 07/02/2014
Event Type  Injury  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.No 510# number provided because this implant is sold internationally with different indications for use; it is currently sold in the us under a different part number.The correction/removal reporting number listed applies to the corresponding product code sold domestically.
 
Event Description
Asr revision, asr resurfacing - left, reason(s) for revision: unknown.Surgeon confirmation form received (b)(6) 2014.Revision reasons added.Revision date amended.Reason(s) for revision: alval / soft tissue reaction, high metal ion level.Update (b)(6) 2016: added patient name, initials, gender and dob.Added additional surgeon.Taken from updated asr legal claim (b)(6) 2016 ((b)(4)).Update (b)(6) 2017: litigation received.Litigation alleges high metal ions.It was also reported that after the revision surgery, patient is still experiencing residual pain, frequent episodes of cramps and muscle weakness.Unknown stem and sleeve have been added.Added complainant information.This was updated on jul 21, 2017.
 
Manufacturer Narrative
Asr revision asr resurfacing - left reason(s) for revision: unknown surgeon confirmation form received (b)(6) 2014.Revision reasons added.Revision date amended.Reason(s) for revision: alval / soft tissue reaction, high metal ion level update (b)(6) 2016: added patient name, initials, gender and dob.Added additional surgeon.Taken from updated asr legal claim (b)(6) 2016 (pd (b)(6) 2016) update (b)(6) 2017: litigation received.Litigation alleges high metal ions.It was also reported that after the revision surgery, patient is still experiencing residual pain, frequent episodes of cramps and muscle weakness.Unknown stem and sleeve have been added.This was updated on (b)(6) 2017.The reported event has been evaluated and will be monitored.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Asr revision asr resurfacing - left reason(s) for revision: unknown surgeon confirmation form received (b)(4) 2014.Revision reasons added.Revision date amended.Reason(s) for revision: alval / soft tissue reaction, high metal ion level update (b)(4) 2016: added patient name, initials, gender and dob.Added aditional surgeon.Taken from updated asr legal claim (b)(4) 2016 (pd (b)(4) 2016) update (b)(4) 2017: litigation received.Litigation alleges high metal ions.It was also reported that after the revision surgery, patient is still experiencing residual pain, frequent episodes of cramps and muscle weakness.Unknown stem and sleeve have been added.This was updated on (b)(4) 2017.Update (b)(4) 2017: litigation record received.It was stated that the chromium-cobalt level were above 7ug/l.Added complainant information.This complaint was updated on (b)(4) 2017.The reported event has been evaluated and will be monitored.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Update oct 10, 2017: litigation record received.It was stated that the chromium-cobalt level were above 7ug/l.Added complainant information.This complaint was updated on oct 19, 2017.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN DEPUY XL TAPER SLEEVE ADAPTOR
Type of Device
HIP FEMORAL STEM/SLEEVE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6734021
MDR Text Key80747648
Report Number1818910-2017-21584
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor,health p
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date 06/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK-HIP
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/22/2017
Initial Date FDA Received07/21/2017
Supplement Dates Manufacturer Received08/10/2017
10/10/2017
11/17/2017
Supplement Dates FDA Received08/14/2017
10/20/2017
11/24/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age52 YR
-
-