The actual device was not available; however, fifteen (15) companion samples were received for evaluation.A batch review was conducted and there were no deviations found related to the reported condition during the manufacture of this lot.A visual inspection was performed with the naked eye.Functional testing was performed which included leak testing, clear passage testing, clamp function testing, and device-device interaction testing.Functional testing passed with no issues.Visual inspection revealed damaged cassette sheeting on one sample; the other 14 samples had no issues.The reported problem was not verified because leak testing was performed with no issues.One sample did not meet specification.Should additional relevant information become available, a supplemental report will be submitted.
|