C.R. BARD, INC. (COVINGTON) -1018233 ROCHESTER MEC ULTRAFLEX M 30BX Back to Search Results
Catalog Number 33302
Device Problems Difficult to Remove (1528); Sticking (1597)
Patient Problem No Information (3190)
Event Date 07/07/2017
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event Description
It was reported that the customer had problems removing the sheaths.The patient stated that the device was "stickier" than usual.
 
Manufacturer Narrative
The reported event was unconfirmed.Received 5 mec lot samples for evaluation.The samples were visually inspected for holes or damage on the package, legibility of printed numbers and letters, expiration date, correct impression, print color, missing parts of printing, position of printing, damaged component, and foreign matter.No issues were found in the samples.The samples were evaluated according to the alleged failure mode: "batch seem stickier".Test results: (b)(6).All samples were within specification.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "description/indication the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply 1) wash penis with mild soap and warmwater.Dry thoroughly.2) trim pubic hair if necessary.3) unroll self-adhering catheter over penis.4) gently squeeze the catheter to properly seal adhesive to the skin.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.5) connect to drainage device.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." (b)(4).
 
Event Description
It was reported that the customer had problems removing the sheaths.The patient stated that the device was "stickier" than usual.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROCHESTER MEC ULTRAFLEX M 30BX
Type of Device
MEC
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
parque industrial aeropuerto
covington GA 30014
MDR Report Key6739404
MDR Text Key81067311
Report Number1018233-2017-03915
Device Sequence Number1
Product Code NNX
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 02/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/25/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/31/2021
Device Catalogue Number33302
Device Lot NumberNGAW0447
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2017
Date Manufacturer Received02/20/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-