|
Catalog Number RBY4C0860 |
Device Problems
Bent (1059); Failure to Advance (2524); Physical Resistance (2578)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/28/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Results: the embolization coil was intact with the ruby coil pusher assembly.There was clotted blood present on the embolization coil.The distal portion of the embolization coil had several offset coil winds.The pusher assembly was kinked approximately 104.0 and 116.0 cm from the proximal end.During functional analysis, the ruby coil was advanced out of its introducer sheath with some difficulty after being flushed with cavicide solution.Conclusions: evaluation of the returned ruby coil revealed that the embolization coil had multiple offset coil windings and the pusher assembly was kinked.If an rhv is not used in the procedure, the introducer sheath may not maintain proper alignment, and resistance may be experienced.The damage observed on the embolization coil can occur if the introducer sheath is not properly aligned inside of the hub of the microcatheter during advancement of the ruby coil.Continued attempts to manipulate the coil can further damage the embolization coil and may contribute to any resistance experienced during the procedure.The pusher assembly kinks were likely a result of continued attempts to advance the embolization coil against resistance.The clotted blood present on the embolization coil is likely a result of not maintaining continuous flush throughout the procedure.The non-penumbra microcatheter mentioned in the complaint and pod8 were not returned for evaluation.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2017-01239.
|
|
Event Description
|
The patient was undergoing a coil embolization procedure in the hepatic artery using ruby coils and pod8''s.During the procedure, the physician successfully deployed and detached initial penumbra coils into the target vessel using a non-penumbra microcatheter.While attempting to advance a new ruby coil through the microcatheter, the physician experienced resistance and subsequently, the ruby coil pusher assembly became bent.Therefore, the ruby coil was removed and it was observed that the microcatheter appeared to be bent.The physician then deployed and detached additional penumbra coils into the target vessel using the same microcatheter.While attempting to pull a new pod8 out of its packaging hoop, the physician inadvertently bent the pod8 pusher assembly.Therefore, the pod8 was set aside and the procedure was successfully completed using four ruby coils, two pod packing coils (podj''s), one pod6 and the same microcatheter.There was no report of an adverse effect to the patient.
|
|
Search Alerts/Recalls
|
|
|