• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARIBEX NOMAD PRO 2; EXTRAORAL SOURCE X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARIBEX NOMAD PRO 2; EXTRAORAL SOURCE X-RAY SYSTEM Back to Search Results
Model Number 0.850.0021
Device Problems Overheating of Device (1437); Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 06/29/2017
Event Type  malfunction  
Manufacturer Narrative
An evaluation has not been conducted as all affected components have not yet been received.It is anticipated that aribex will receive the affected components and an evaluation will be conducted.A follow-up report will be submitted upon evaluation completion.
 
Event Description
It was reported that the charging cradle has melted and is no longer working.The battery was not on the cradle at the time.There was no report of injury, user or patient involvement, or impact to patient care.
 
Manufacturer Narrative
The unit was returned and an evaluation was completed.Upon visual inspection it appears that a thermal event occurred.It is not possible to determine the exact sequence of events that led to the thermal event.However, there is evidence that a short occured in the lower battery pack between the cells and printed circuit board.There was a blackening on the top of the lower battery pack and its printed circuit board.Cells 2 and 3 appear to have been involved in the thermal event, likely by supplying energy into a fault in the board.Additionally there was a slight melting of the handset plastic enclosure.This concludes the investigation.A recall was conducted and termination of the recall has been requested.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOMAD PRO 2
Type of Device
EXTRAORAL SOURCE X-RAY SYSTEM
Manufacturer (Section D)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer (Section G)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer Contact
erika martin
11727 fruehauf drive
charlotte, NC 28273
7045877241
MDR Report Key6749882
MDR Text Key81245471
Report Number1017522-2017-00045
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081664
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 06/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0.850.0021
Device Catalogue Number0.850.0021
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/29/2017
Initial Date FDA Received07/28/2017
Supplement Dates Manufacturer Received06/29/2017
Supplement Dates FDA Received08/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-