Model Number LN130B
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The sample was not returned for evaluation; therefore, the complaint was not confirmed and a root cause was not able to be determined.From the complaint description, the issue seems to be with the lack of slit in the duckbill valve, which would then not allow flow through the valve during use.During assembly of ops valves, all duckbill valves are to be squeezed and confirmed to open prior to placing them into the housing.Ops valves are also 100% leak tested in process, during which a flow test is performed to confirm the duckbill opens and allows flow in the correct direction.This issue is currently undergoing an investigation; however, it is presently believed that the lack of flow through the duckbill valve is due to a process change in the water that the duckbills are washed in at the supplier.(b)(4).All available information has been placed on file in quality management for appropriate tracking, trending and follow-up.
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during setup, the ops valves in the tubing pack would not open.*no consequences or impact to patient, *product was not changed out.*procedure was completed successfully.
 
Manufacturer Narrative
(b)(4).The investigation results and conclusions, previously reported, remain the same.
 
Event Description
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALVE, O.P.S BULK, N-S
Type of Device
OVERPRESSURE SAFETY VALVE
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer Contact
cathleen hargreaves
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key6749937
MDR Text Key81249908
Report Number1124841-2017-00161
Device Sequence Number1
Product Code DTL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K820297
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 12/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2019
Device Model NumberLN130B
Device Catalogue NumberN/A
Device Lot NumberUH25
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/25/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1124841-06/25/2017-001-R
Patient Sequence Number1
-
-